CLIA Kit for Triiodothyronine (T3) - UCCA453Ge

CLIA Kit for Triiodothyronine (T3) - UCCA453Ge

Regular price
$571.20
Regular price
Sale price
$571.20
Unit price
per 
Availability
Sold out

CLIA Kit for Triiodothyronine (T3)

Size: 96T

Catalogue Number: UCCA453Ge-96

Citations, Manuals and MSDS Available upon request.

Target: Triiodothyronine (T3)

Alternative Names: 3,3',5-Triiodo-L-Thyronine; Liothyronine; Cytomel; Tertroxin

Target Species: General species

Detection range: 78.1-20,000pg/mL

Sensitivity: 25.9pg/mL

Featured Series Function: Detects small molecule

Recommended/Predicted Sample Types: Serum, Plasma and other Biological Fluids

Assay Time: 2h

Method: Chemiluminescence

Specificity: Reactive with General species T3 / Triiodothyronine

Detection principle: Competitive Inhibition

Sample Size: 50uL

Assay Precision: Intra-Assay: CV<10%, Inter-Assay: CV<12%

Storage: 4°C for 1 month/ -20°C for long-term(One year within shelf life)

Shelf-life: 12 months

Specificity: This assay has high sensitivity and excellent specificity for detection of Triiodothyronine (T3).
No significant cross-reactivity or interference between Triiodothyronine (T3) and analogues was observed.

Precision: Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Triiodothyronine (T3) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Triiodothyronine (T3) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100

Stability: The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary: 1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
And then add 50µL prepared Detection Reagent A immediately.
Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
7. Read RLU value immediately.

Test principle: The microplate provided in this kit has been pre-coated with a monoclonal antibody specific to Triiodothyronine (T3). A competitive inhibition reaction is launched between biotin labeled Triiodothyronine (T3) and unlabeled Triiodothyronine (T3) (Standards or samples) with the pre-coated antibody specific to Triiodothyronine (T3). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Triiodothyronine (T3) in the sample. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is reverse proportional to the Triiodothyronine (T3) level in the sample or standard.

Research Area: Endocrinology; Autoimmunity;

Research Use Only